Clinical Observation of ICL Diameter Selection by Using Sirius Anterior Segment Analysis System |
Wenhui Zeng1 , Dongqiang Luo2 , Dingjuan Zhong2 , Hua Wang2 |
1 Department of Ophthalmology, People's Hospital of Changsha County, Hunan Provincial People's Hospital Xingsha Branch, Changsha 410100, China 2 Center of Ophthalmology and Optometric, Hunan Provincial People's Hospital, the First Affiliated Hospital of Hunan Normal University, Changsha 410000, China |
|
|
Guide |
|
Abstract Objective: To evaluate the reference value of selecting implantable collamer lens (ICL) size by Sirius anterior segment analysis system. Methods: We retrospectively collected 146 eyes of 73 patients with high myopia who underwent ICL implantation in Hunan Provincial People's Hospital from January 2016 to August 2017. All ICL diameters were selected by the anterior segment indicators such as horizontal visible iris diameter (HVID), horizontal anterior chamber diameter (HACD), anterior chamber depth (ACD), which was measured by Sirius anterior segment analysis system. ICL calculation software provided by STAAR company is used to select ICL diameter based on these indicators. All patients were followed 3 months to 1 year. The data before and after surgery were analyzed using Friedman M test, repeated measures ANOVA, paired t test and multiple linear regression equation. Results: The postoperative uncorrected distance visual acuity (UDVA) of all patients was improved compare with preoperative value (χ2 =156.08, P<0.001). The effective index and safety index were greater than 1 at each postoperative period. There was no significant difference in intraocular pressure (IOP) during the preoperative and postoperative follow-up (P=0.054). The endothelial cell density (ECD), anterior chamber depth (ACD) and anterior chamber angle (ACA) were decreased after surgery (t=4.63, 31.17, 23.57, all P<0.001). The postoperative mean vault was (483±231) μm and the ideal vault eyes accounted for 83.6%. ICL diameter (mm)=6.474+0.43×HVID (mm) +0.299×ACD (mm) (R2 =0.44, F=36.23, P<0.001). No complications such as cataract, goniosynechia and glaucoma occurred in all patients. Conclusions: The ICL selected by Sirius anterior segment analysis system is safe and effective in the treatment of high myopia; the HVID, HACD, and ACD obtained by Sirius can be used as important parameters for ICL diameter selection.
|
Received: 12 September 2021
|
|
Fund:Special Task of Hunan Innovative Province Construction Project (2020SK2118) |
Corresponding Authors:
Hua Wang, Center of Ophthalmology and Optometric, Hunan Provincial People's Hospital, the First Affiliated Hospital of Hunan Normal University, Changsha 410000, China(Email:wanghuaeye@163.com)
|
|
|
|
[1] |
罗栋强, 王华, 陶思思, 等. 有晶体眼后房型人工晶体植入 术矫正超高度近视的安全性. 中华眼视光学与视觉科学 杂志, 2014, 12(16): 717-721. DOI: 10.3760/cma.j.issn.1674- 845X.2014.12.003.
|
[2] |
崔同峰, 周进, 王铮. ICL植入术后异常拱高相关影响因素. 中华眼视光学与视觉科学杂志, 2019, 21(7): 534-539. DOI: 10.3760/cma.j.issn.1674-845X.2019.07.009.
|
[3] |
陶思思, 王华, 刘鹏飞, 等. 水平角膜直径和前房直径与睫状 沟直径的相关性研究. 国际眼科杂志, 2019, 19(6): 1031-1034. DOI: 10.3760/cma.j.issn.1674-845X.2019.07.009.
|
[4] |
Kumar M, Shetty R, Jayadev C, et al. Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes. Indian J Ophthalmol, 2017, 65(4): 288-294. DOI: 10.4103/ijo.IJO_729_16.
|
[5] |
胡宗莉 , 叶 剑 . 有晶状体眼后房型人工晶状体植入矫 正屈光不正的临床进展 . 中华眼视光学与视觉科学杂 志, 2019, 21(9): 715-720. DOI: 10.3760/cma.j.issn.1674- 845X.2019.09.012.
|
[6] |
Elmohamady MN, Abdelghaffar W. Anterior chamber changes after implantable collamer lens implantation in high myopia using pentacam: a prospective study. Ophthalmol Ther, 2017, 6(2): 343-349. DOI: 10.1007/s40123-017-0109-3.
|
[7] |
Eissa SA, Sadek SH, El-Deeb MW. Anterior chamber angle evaluation following phakic posterior chamber collamer lens with CentraFLOW and its correlation with ICL vault and intraocular pressure. J Ophthalmol, 2016, 2016: 1383289. DOI: 10.1155/2016/1383289.
|
[8] |
Packer M. Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens. Clin Ophthalmol, 2016, 10: 1059-1077. DOI: 10.2147/OPTH. S111620.
|
[9] |
Alfonso JF, Fernández-Vega L, Lisa C, et al.Long-term evaluation of the central vault after phakic Collamer lens (ICL) implantation using OCT. Graefes Arch Clin Exp Ophthalmol, 2012, 250(12): 1807-1812. DOI: 10.1007/s00417-012-1957-0.
|
[10] |
Packer M. The Implantable Collamer Lens with a central port: review of the literature. Clin Ophthalmol, 2018, 12: 2427-2438. DOI: 10.2147/OPTH.S188785.
|
[11] |
Montés-Micó R, Ruiz-Mesa R, Rodríguez-Prats JL, et al. Posterior-chamber phakic implantable collamer lenses with a central port: a review. Acta Ophthalmol, 2021, 99(3): e288-e301. DOI: 10.1111/aos.14599.
|
[12] |
Benda F, Filipova L, Filipec M. Correction of moderate to high hyperopia with an implantable collamer lens: medium- term results. J Refract Surg, 2014, 30(8): 526-533. DOI: 10.3928/1081597X-20140711-05.
|
[13] |
Kocová H, Vlková E, Michalcová L, et al. Implantation of posterior chamber phakic intraocular lens for myopia and hyperopia-long-term clinical outcomes. J Fr Ophthalmol, 2017, 40(3): 215-223. DOI: 10.1016/j.jfo.2016.10.009.
|
[14] |
Reinstein DZ, Lovisolo CF, Archer TJ, et al.Comparison of postoperative vault height predictability using white-to-white or sulcus diameter-based sizing for the visian implantable collamer lens. J Refract Surg, 2013, 29(1): 30-35. DOI: 10.3928/1081597X-20121210-02.
|
[15] |
Seo JH, Kim MK, Wee WR, et al. Effects of white-to-white diameter and anterior chamber depth on implantable collamer lens vault and visual outcome. J Refract Surg, 2009, 25(8): 730- 738. DOI: 10.3928/1081597X-20090707-08.
|
|
|
|