Objective Compare the consistency in calculating the IOL power after LASIK surgery between the Sirius Ray-Tracing method and IOLMaster. Methods In the prospective clinical study, 89 eyes of 89 patients who underwent myopic and myopic astigmatism LASIK surgery were selected according to preoperative SE and divided into two groups, 59 eyes for mild and moderate myopia(≤-6.0 D), 30 eyes for high myopia(>-6.0 D). Haigis formula was used to measure the IOL(The types of IOL are Alcon SN60WF and AMO Tecnis ZA9003) power before LASIK, and Haigis-L formula, Haigis formula were used after LASIK 1 month by IOLMaster, and compare the results with Sirius Ray-Tracing method. The Bland-Altman was used to analyze the results. Results The comparison of Sirius method and Haigis formula preoperative: Bland-Altman analysis showed the 95%LoA of mild and moderate myopia group of two type IOL were(-0.78~+0.84)D, (-0.57~+1.21)D, respectively. There were 3%, 3% of the points beyond their respective 95%LoA. The intraclass correlation coefficient (ICC) were 0.976 and 0.945, respectively. The 95%LoA of high myopia group were (-0.83~+0.69)D, (-0.57~+0.93)D. There were 3%, 3% of the points beyond their respective 95%LoA. The ICC were 0.981 and 0.979, respectively. The comparison of Sirius method and Haigis formula postoperative: the 95%LoA of mild and moderate myopia group of two type IOL were (-0.27~+2.59)D, (-0.50~+3.30)D. There were 5%, 5% of the points beyond their respective 95%LoA. The ICC were 0.641, 0.584, respectively. The 95%LoA of high myopia group were (-0.48~+2.78)D, (-0.17~+3.17)D. There were 10%, 7% of the points beyond their respective 95%LoA. The ICC were 0.622, 0.651, respectively. The comparison of Sirius method and Haigis-L formula postoperative: the 95%LoA of mild and moderate myopia group were (-1.02~+0.58)D, (-1.10~+0.77)D. There were 0, 2% of the points beyond their respective 95%LoA. The ICC were 0.986 and 0.955, respectively. The 95%LoA of high myopia group were (-2.36~+0.80)D, (-1.98~+1.22)D. There were 7%, 10% of the points beyond their respective 95%LoA. The ICC were 0.667 and 0.638, respectively. Conclusion There was better consistency of Sirius Ray-Tracing method and Haigis formula in unoperative eyes, and better consistency with Haigis-L formula in mild and moderate myopia after LASIK, but poor consistency with Haigis-L formula in high myopia after LASIK to calculate the IOL power.
杨应琴,王卫群,田英,刘欢,马艳丽. LASIK术后Sirius光线追踪法与IOLMaster法测量人工晶状体度数比较[J]. 中华眼视光学与视觉科学杂志, 2016, 18(6): 363-372.
Yang Yingqin,Wang Weiqun,Tian Ying,Liu Huan,Ma Yanli. Comparison of Sirius ray-tracing method and IOLMaster in calculating the intraocular lens power after LASIK . Chinese Journal of Optometry Ophthalmology and Visual science, 2016, 18(6): 363-372.
Saiki M, Negishi K, Kato N,et al.Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery[J]. Jpn J Ophthalmol,2014,58(3):276-281. DOI:10.1007/s10384-014-0304-x.
[5]
Hoffer KJ. Intraocular lens power calculation after previous laser refractive surgery[J]. J Cataract Refract Surg,2009,35(4):759-765. DOI:10.1016/j.jcrs.2009.01.005.
[6]
Haigis W, Lege B, Miller N, et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis[J]. Graefes Arch Clin Exp Ophthalmol,2000,238(9):765-773.
[7]
Haigis W. Intraocular lens calculation after refractive surgery for myopia: Haigis-L formula[J]. J Cataract Refract Surg,2008,34(10): 1658-1663. DOI:10.1016/j.jcrs.2008.06.029.
[11] Savini G, Barboni P, Carbonelli M, et al. Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography[J]. J Cataract Refract Surg,2011,37(10): 1809-1816. DOI:10.1016/j.jcrs.2011.04.033.
[12] Savini G, Barboni P, Carbonelli M, et al. Accuracy of corneal power measurements by a new Scheimpflug camera combined with Placido-disk corneal topography for intraocular lens power calculation in unoperated eyes[J]. J Cataract Refract Surg,2012, 38(5):787-792. DOI:10.1016/j.jcrs.2011.11.037.
[13] Savini G, Bedei A, Barboni P, et al.Intraocular lens power calculation by ray-tracing after myopic excimer laser surgery[J].Am J Ophthalmol,2014,157(1):150-153.e1. DOI:10.1016/j.ajo.2013. 08.006.
[14] Preussner PR, WahI J, Lahdo H, et al. Ray tracing for intraocular lens calculation[J]. J Cataract Refract Surg,2002,28(8):1412-1419.
[15] Miyata K, Otani S, Honbou N, et al. Use of Scheimpflug corneal anterior-posterior imaging in ray-tracing intraocular lens power calculation[J]. Acta Ophthalmol,2013,91(7):e546-e549. DOI:10.1111/aos.12139.