Validity of Presbyopic Addition Lenses Based on Clouclip |
Dian Zhou1, 2, Xiaoning Li2, 3, Lun Pan2 , Haoran Wu1, 2, Hong Zhong2 , Zhikuan Yang1, 2, 3 |
1 Aier School of Ophthalmology, Central South University, Changsha 410000, China 2 Aier Institute of Optometry and Vision Science, Changsha 410000, China 3 Aier School of Optometry, Hubei University of Science and Technology, Xianning 437000, China |
|
|
Guide |
|
Abstract Objective: To evaluate the validity of the presbyopic prescription based on Clouclip (Glasson Technology Co., Ltd, Hangzhou, China). Methods: A total of 181 subjects aged 28-60 years old with normal BCVA were included in the cross-sectional study. Refractive errors were corrected using frame glasses. Clouclip was used to monitor the habitual reading distance of subjects and the near addition (ADD) was calculated with habitual reading distance. The ADD was tested by both the traditional method and Clouclip. The habitual reading distance was tested first, then the ADD of Clouclip was calculated based on habitual reading distance. Subjects were divided into a pre-presbyopia group (<40 years old) and a presbyopia group (≥40 years old) according to age. Subjects also were divided into a myopia group (<-0.50 D) and emmetropia group (-0.50- +0.50 D) based on refractive error. A factorial design ANOVA was used to analyze the influence of age and refractive error on habitual reading distance. A paired t-test, association analysis and Bland-Altman analysis were used to evaluate the consistency of traditional ADD and the ADD of Clouclip. Results: In all subjects, the presbyopic group had a farther reading distance than the pre-presbyopic group (0.37±0.06 m vs. 0.38±0.06 m, F=5.51, P=0.02), and the habitual reading distance was not affected by refractive status (F=1.48, P=0.23). The ADD of Clouclip was highly correlated with the traditional ADD (r=0.74, P<0.001). The Bland-Altman analysis showed that the average deviation was only -0.08 D. Conclusions: The habitual reading distance is affected by presbyopic status. The ADD of Clouclip is highly consistent with the traditional ADD. The method of using Clouclip to calculate ADD is an objective, accurate and personalized method for presbyopic prescriptions.
|
Received: 15 April 2019
|
|
Fund: Aier Eye Hospital Group Research Project (AF1603D2, AM163D06) |
Corresponding Authors:
Zhikuan Yang, Aier School of Ophthalmology, Central South University, Changsha 410000, China; Aier Institute of Optometry and Vision Science, Changsha 410000, China; Aier School of Optometry, Hubei University of Science and Technology, Xianning 437000, China (Email: 13380071988@ 189.cn)
|
|
|
|
[1] |
郑荣领, 金秀英. Monovision矫正老视研究进展. 眼视光学杂志, 2000, 2(1): 59-61. DOI: 10.3760/cma.j.issn.1674- 845X.2000.01.020.
|
[2] |
Holden BA, Fricke TR, Ho SM, et al. Global vision impairment due to uncorrected presbyopia. Arch Ophthalmol, 2008, 126(12): 1731-1739. DOI: 10.1086/379289.
|
[3] |
施青, 毕燕龙. 老视治疗的研究进展. 同济大学学报(医学版), 2015, 36(6): 128-132. DOI: 10.16118/j.1008-0392.2015.06.026.
|
[4] |
杨智宽. 临床视光学. 2版, 北京: 科学出版社, 2014: 190-191.
|
[5] |
温龙波, 蓝卫忠, 李晓柠. 客观监测近视眼相关环境因素的新设备“云夹”的准确性和稳定性研究. 中华眼视光学与视觉科学杂志, 2017, 19(4): 198-203. DOI: 10.3760/cma.j.issn. 1674-845X.2017.04.002.
|
[6] |
Leung TW, Flitcroft DI, Wallman J, et al. A novel instrument for logging nearwork distance. Ophthalmic Physiol Opt, 2011, 31(2): 137-144. DOI: 10.1111/j.1475-1313.2010.00814.x.
|
[7] |
瞿佳. 视光学理论和方法. 北京人民卫生出版社, 2004: 169- 163.
|
[8] |
Drobe B, Lombo I. Working distance in non-presbyopes and progressive lens wearing presbyopes. Inves Ophthalmol Vis Sci, 2006, 47(13): 5861-5861.
|
[9] |
Boring, Edwin G. Helmholtz's treatise on physiological optics. Psychol Bull, 1926, 23(6): 323-326. DOI: 10.1037/h0066480.
|
[10] |
Saw SM, Foster PJ, Gazzard G, et al. Undercorrected refractive error in Singaporean Chinese adults: The Tanjong pagar survey. Ophthalmology, 2004, 111(12): 2168-2174. DOI: 10.1016/ j.ophtha.2004.05.032.
|
[11] |
Atchison DA. New thinking about presbyopia. Clin Exp Optom, 2008, 91(3): 205-206. DOI: 10.1111/j.444-0938.2008.00264.x.
|
[12] |
赵瑾, 谢毅杰, 马丽卿. 屈光不正与年龄、性别、眼别关系的研究. 中国实用眼科杂志, 1992, 10(7): 407-409.
|
|
|
|