Abstract:Objective: To explore the optical coherence tomography (OCT) characteristics of foldable capsular vitreous body (FCVB) implanted in eyes with severe retinal detachment. Methods: Serial case study. 10 cases (10 eyes) with severe retinal detachment were selected for FCVB implantation in Beijing Huaer Hospital from November 2017 to March 2021. Preoperative and postoperative fundus examinations were performed by OCT in 1 week, 6 months, 1 year, and 2 years. To analyze retinal reattachment after FCVB implantation by comparing the preoperative and postoperative results of fundus examinations. Results: Fundus photography showed that the fundus was obviously visible in 9 out of 10 eyes with FCVB implantation, and was lack of clarity in 1 eye due to mild corneal opacity. Postoperative OCT examination was found that a high-reflective optical strip with a thickness of 200 μm in front of the retina was in 10 eyes, the macular fovea structure was normal and the retina was basically reattached was in 9 eyes, and no retinal signal was found in 1 eye. At the same time, it was found that the gap between the FCVB capsule and the retina was small, and the retina was well attached by FCVB. Conclusions: After FCVB implantation, there is a characteristic high-reflective optical band in front of the retina, which is obviously different from the vitreous cavity and the retina. At the same time, the reattachment of the retina can be observed.
宋云红, 王绍莉. 折叠式人工玻璃体球囊植入严重视网膜脱离眼内的OCT特征[J]. 中华眼视光学与视觉科学杂志, 2022, 24(1): 52-57.
Yunhong Song,Shaoli Wang. The OCT Imaging Characteristics of Foldable Capsular Vitreous Body Implanted in Eyes with Severe Retinal Detachment. Chinese Journal of Optometry Ophthalmology and Visual science, 2022, 24(1): 52-57. DOI: 10.3760/cma.j.cn115909-20210528-00222
Gao Q, Mou S, Ge J, et al. A new strategy to replace the natural vitreous by a novel capsular artificial vitreous body with pressurecontrol valve. Eye, 2008, 22(3): 461-468. DOI: 10.1038/sj.eye.6702875. Epub 2007 May 25.
[2]
Liu Y, Jiang Z, Gao Q, et al. Technical standards of foldablecapsular vitreous body regarding mechanical, optical and biocompatible properties. Artif Organs, 2010, 34(10): 836-845.DOI: 10.1111/j.1525-1594.2010.01006.x.
[3]
Lin X, Ge J, Gao Q, et al. Evaluation of the flexibility, efficacy,and safety of a foldable capsular vitreous body in the treatment of severe retinal detachment. Invest Ophthalmol Vis Sci, 2011,52(1): 374-381. DOI: 10.1167/iovs.10-5869.
[4]
Chen J, Gao Q, Liu Y, et al. Clinical device-related article evaluation of morphology and functions of a foldable capsular vitreous body in the rabbit eye. J Biomed Mater Res B Appl Biomater, 2011, 97(2): 396-404. DOI: 10.1002/jbm.b.31812.Epub 2011 Mar 25.
[5]
Chen S, Tian M, Zhang L, et al. Reattachment after foldable capsular vitreous body implantation in severe retinal detachment eyes. Transl Vis Sci Technol, 2021, 10(11): 8. DOI: 10.1167/tvst.10.11.8.
[6]
Lin X, Sun X, Wang Z, et al. Three-year efficacy and safety of a silicone oil-filled foldable-capsular-vitreous-body in three cases of severe retinal detachment. Transl Vis Sci Technol, 2016, 5(1):2. DOI: 10.1167/tvst.5.1.2.
[7]
Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science, 1991, 254(5035): 1178-1181. DOI:10.1126/science.1957169.
Zhang X, Tian X, Zhang B, et al. Study on the effectiveness and safety of foldable capsular vitreous body implantation. BMC Ophthalmol, 2019, 19(1): 260. DOI: 10.1186/s12886-019-1268-x.
[10]
Wang P, Gao Q, Jiang Z, et al. Biocompatibility and retinal support of a foldable capsular vitreous body injected with saline or silicone oil implanted in rabbit eyes. Clin Exp Ophthalmol,2012, 40(1): e67-75. DOI: 10.1111/j.1442-9071.2011.02664.x.